CMOS Tubing Inspection Cabinet (TIC) VSA 4/2003
|
|
|
Over twice the resolution of Amorphous Silicon and and much less noise
|
|
![]() |
The CMOS tubing inspection cabinet, is the highest resolution turn-key solution for small tubing inspection applications. The mobile shielded cabinet based system incorporates a computer controlled micro focus X-ray tube, a high-resolution CMOS digital X-ray detector, and a rotational movement system for capturing images between 0-120 degrees.
At the comfortable operator workstation, an intelligent interface allows the operator to drastically reduce the inspection times by simultaneously acquiring digital images, and automatically repositioning detector to pre-assigned positions around tubing cross section. On the spot interpretations can be preformed and efficient image archiving can be obtained to meet inspection requirements. |
|
|
|
| image area: | 2 x 2 inches |
|
Operational information: Image acquisition takes less than 10 seconds from "shot to screen", and positional relocation happens immediately after X-ray is obtained. Full movement 0-120º (without interruption) requires approximately 8 seconds.
Tubing is easily placed/removed in cabinet either through top access port or by direct slide through entry. Safety interlocks and X-ray on indicators assure safe operation. The TIC cabinet meet US safety and radiation standards (ANS/NBS CFR Sect. 1020.40) for full protection installation devices. |
|
| Response: | 20 - 160 kV |
| spatial resolution: | 50 micron, 10 lp/mm |
| Dynamic Range: | 12 bit (4096 grayscale) |
| Speed: | less than 10 seconds / image |
| software: | EnSpect image capture package with expanded interpretation tools and image archiving packet, Incorporated image/source manipulation control. |
| Presentation: | Images can be magnified, regional contrast-enhancement, colorization, measurement, invert, emboss, annotation, spot magnification, histogram, and other tools. |
|
Data Archiving:
|
Images can be saved on CD-ROM or DVD as, .bmp, .tif, and Dicom |
|
Specifications subject to change without notice
|
|